ISOL Technologies MRI Systems Comparison Chart

ModelRelaxChorusEncoreForte
Clinical ApplicationWhole bodyWhole bodyWhole bodyWhole body
ConfigurationOpen MRIShort BoreShort Bore CompactShort Bore Compact
SynchronizationECG/peripheral; Optional/yes, respiratory gatingECG/peripheral; Optional/yes, respiratory gatingECG/peripheral; Optional/yes, respiratory gatingECG/peripheral; Optional/yes, external trigger, respiratory gating
Imaging Modes3D volume imaging (MIP, MPR) MR Angiography package2D/3D, Travelling Sat, Multi-Slab 3D, MTC and TONE Pulse Sequence, Fat/Water Suppression, Presaturation (up to 6 bands), Flow Compensation using GMR pulse, Multi-Slice, Multi-Group Imaging2D/3D, Travelling Sat, Multi-Slab 3D, MTC and TONE Pulse Sequence, Fat/Water Suppression, Presaturation (up to 6 bands), Flow Compensation using GMR pulse, Multi-Slice, Multi-Group Imaging2D/3D, T1, T2 and Diffusion//Perfusion imaging, MR Angiography package, Advanced EPI package, Multi-nuclei MR Spectroscopy package
FOV30 mm40 cm40 cm40 cm
Magnet TypePermanentSuperconductingSuperconductingSuperconducting
Bore Diameter40 cm diameter (patient)58 cm diameter (patient)58 cm diameter61 cm without body coil
Magnet Weight13500 kg4050 kg3200 kg12000 kg
H*W*D170 x 210 x 130 cm233 x 206 x 160 cm200 x 168 x 187 cm260 x 220 x 235 cm
Power RequirementsN/AN/AN/AN/A
Cooling SystemWater-cooled coil and air-cooled amplifierWater-cooled coil and air-cooled amplifierAir-cooled coil and amplifierWater-cooled coil and air-cooled amplifier
Cryogen UseN/A0.07 L/hr helium0.05 L/hr helium0.15 L/hr helium
Field Strength0.35 T1.5 T0.5 T3.0 T
Strength25 – 30 mT/m20 mT/m (up to 27 mT/m)15 mT/m38 mT/m
5 Guass Fringe Fieldlower than 2.4 m from the iso-center2.5 m/3.8 m2.3 m/3.1 m3.3 m/5.2 m
Back To Top