GE Open Bore MRI Systems Comparison Chart

ModelSigna OpenSpeedSigna OvationSigna ProfileSigna 3.0 T
Clinical ApplicationWhole bodyWhole bodyWhole bodyWhole body
ConfigurationOpen MRIOpen MRIOpen MRICylindrical – high homogeneity
SynchronizationStandard cardiac gating, ECG/peripheral, respiratory gatingStandard cardiac gating, ECG/peripheral, respiratory gatingStandard cardiac gating, ECG/peripheral, respiratory gatingECG/peripheral, respiratory gating
Imaging ModesLocalizer, single, multislice, volume, fast, POMP, multi slab, cineLocalizer, single, multislice, volume, fast, POMP, multi slab, cine, slice and frequency zip, extended dynamic range, tailored RFLocalizer, single, multislice, volume, fast, POMP, multi slab, cine, slice and frequency zip, extended dynamic range, tailored RFSingle, multislice, volume, fast, POMP, multi slab, cine
FOV1 cm to 40 cm continuous3 cm to 40 cm continuous3 cm to 40 cm continuous1 cm to 40 cm continuous
Magnet TypeSuperconductingPermanentPermanentSuperconducting
Bore Diameter175 x 85 x 447 cm175 x 85 x 447 cm120 x 44 cm55 cm
Magnet Weight8256 kg19200 kg10,000 kg w/ gradient enclosure15102 kg incl. cryogen’s
H*W*D530 x 175 x 250 cm530 x 175 x 250 cm147 x 214 x 193 cm260 x 238 x 265 cm
Power Requirements200 – 480, 3-phase200 – 480, 3-phase200 – 480, 3-phase480 or 380/415, 3-phase II
Cooling SystemLiquid heliumN/ANone requiredClosed-loop water-cooled grad.
Cryogen Use0.3 L/hr, holds 300 LN/AN/ALess than 0.14 L/hr liquid He
Field Strength0.7 T0.35 T0.2 T3.0 T
Strength25 mT/mN/A15 mT/m40 mT/m
5 Guass Fringe Field4.7 m/ 3.8 m2.5 m/2.5 m1.7 m/1.7 m5.4 m x 3.2 m
Back To Top