Model | Brilliance iCT | Brilliance 16-slice | MX 16-slice | Ingenuity CT | Brilliance 64-channel | Brilliance Big Bore Oncology |
Detectors, type | Solid-state GOS | Solid-state GOS | 16,128 elements: solid-state GOS | Solid-state GOS with BIP | HiLight ceramic matrix II | Solid-state GOS |
No. of rows | 64 - 128 spherical detectors with 2Danti-scatter | 16 | 16 | 64 (128) | 64 | 16 |
Gantry dimensions, H x W x D, cm | 221 x 89 x 189 | 203 x 239 x 94 | 189 x 221 x 89 | 203 x 238 x 94 | 203 x 238 x 93 | 200 x 252 x 97 |
Gantry weight, kg | 2570 | 1764 | 1850 | 1941 | 1941 | 2043 |
Gantry aperture, cm | 70 | 70 | 70 | 70 | 70 | 85 |
Heat storage, HU | 30,000,000 eff. | 8 (MRC technology) | 5 | MRC Ice (30 MHU eff.) | MRC 8 MHU (26 MHU eff.) | MRC 8 MHU (26 MHU eff.) |
Tube Cooling | Air | Oil/Air | Oil/Air | Air | Air | Air |
kW output | 100-120 | 60 | 50 | 80 | 60 | 60 |
mA Range | 10 to 833 / 1000 | 20 to 500 in 1mA increments | 20 to 420 in 1 mA increments | 20 to 667 in 1mA increments | 20 to 500 in 1mA increments | 20 to 500 in 1mA increments |
Max scan volume, cm | 175 | 175 | 150 | 185 (max. table travel) | 175 (max. table travel) | 175 (max. table travel) |
Max load capacity with accuracy, kg | 204 (295 optional) | 204 +/- 0.25 | 200 +/- 0.25 | 204 (295) | 204 (295) | 204 (295) |
Scan FOVs, cm | 5 to 50 reconstructed FOV | Up to 50 | Up to 50 | 50, 25 | 50, 25 | 60 |
Reconstruction matrices | 512, 768, 1024 | 512 x 512, 768 x 768, 1024 x 1024 | 512 x 512, 768 x 768, 1024 x 1024 | 5,122; 7,682; 10,242 | 5,122; 7,682; 10,242 | 5,122; 7,682; 10,242 |
Dose-modulation technique | DDOM, ZDOM, ASC | DoseRight ACS (automatic current selection), DoseRight D-DOM (rotational dose modulation), DoseRight Z-DOM (longitudinal dose modulation) | DoseRight ACS (automatic current selection), DoseRight DOM (rotational and longitudinal dose modulation) | Yes, angular and longitudinal (Z-DOM) dose modulation techniques | Yes, angular and longitudinal (Z-DOM) dose modulation techniques | Yes, angular (D-DOM) and longitudinal (Z-DOM) dose modulation techniques |
Suggested room size (m2) | 25 | N/S | N/S | 26 (36 optimal) | 26 (36 optimal) | 26 (36 optimal) |