Philips Short Bore MRI Systems Comparison Chart

Model Intera 1.5 T Infinion 1.5 T Evolution 1.5T Omega Evolution 1.5T Omega HP Ingenia 3.0T Ambition 1.5T S Ambition 1.5T X Elition 3.0T S 770 5300 Magnetom Free.Max Magnetom Altea Magnetom Sola Magnetom Ilumina Magnetom Vida Magnetom Prisma
Clinical Application Whole body Whole body Whole body Whole body Whole body Whole body Whole body Whole body Whole body Whole body Whole body Whole body Whole body Whole body Whole body Whole body
Configuration Short Bore Compact Ultra Short Bore Wide bore Wide bore Wide bore Wide bore Wide bore Wide bore Open Bore Wide bore Wide bore Open Bore Open Bore Open Bore Open Bore Open Bore
Synchronization ECG/peripheral: Optional/yes, respiratory gating ECG/peripheral, respiratory gating Respiratory gating
Imaging Modes Single Slice 2D, Multi Single Slice 2D, Multi Slice 2D, 3D, Multi Chunk 3D, Multi Stack 3D Single Slice, Single Volume, Multi Slice, Multi volume MultiSlice
FOV Max. 53 cm 0.4 - 56 cm Max. 55cm Max. 55cm Max. 55cm Max. 55cm Max. 55cm Max. 55cm
Magnet Type Superconducting Superconducting Superconducting Superconducting Superconducting Superconducting Superconducting Superconducting Superconducting Superconducting Superconducting Superconducting Superconducting
Bore Diameter 60 cm diameter (patient) 60 cm diameter (patient) 70 at isocenter, flares to 95 cm 70 at isocenter, flares to 95 cm 70 at isocenter, flares to 95 cm 70 at isocenter, flares to 95 cm 70 at isocenter, flares to 95 cm 70 at isocenter, flares to 95 cm 70 cm 70 at isocenter, flares to 95 cm
Magnet Weight 2900 kg 4100 kg w/ cryogen's 3,400 kg 4,204 kg 4,204 kg 7,350 kg 7,350 kg 9,674 kg
H*W*D 240 x 188 x 157 cm 233 (lead fitted) x 198 x 140 cm 229 x 188 x 150 cm 229 x 188 x 150 cm 229 x 188 x 162 cm 230 x 188 x 162 cm 230 x 188 x 162 cm 230 x 188 x 162 cm 175 x 230 x 207 175 x 230 x 207 145 x 234 x 215 146 x 234 x 215 186 x 245 x 215 186 x 245 x 215 215 x 250 x 231
Power Requirements 380/400 V 400/480 V 400/480 3 phase 400/480 3 phase 400/480 3 phase 400/480 3 phase 400/480 3 phase 380 / 400 / 440 / 480 V, 3 phase + ground 380 / 400 / 420 / 440 / 460 / 480 V, 3 phase + ground 380 / 400 / 420 / 440 / 460 / 480 V, 3 phase + ground 380 / 400 / 420 / 440 / 460 / 480 V, 3 phase + ground 380 / 400 / 420 / 440 / 460 / 480 V, 3 phase + ground 380 / 400 / 420 / 440 / 460 / 480 V, 3 phase + ground
Cooling System N/A Closed-loop, chilled water Direct liquid cooling gradient coil with hollow conductors for maximum cooling power and optimum thermal stability Direct liquid cooling gradient coil with hollow conductors for maximum cooling power and optimum thermal stability Direct liquid cooling gradient coil with hollow conductors for maximum cooling power and optimum thermal stability Direct liquid cooling gradient coil with hollow conductors for maximum cooling power and optimum thermal stability Direct liquid cooling gradient coil with hollow conductors for maximum cooling power and optimum thermal stability Direct liquid cooling gradient coil with hollow conductors for maximum cooling power and optimum thermal stability Direct liquid cooling gradient coil with hollow conductors for maximum cooling power and optimum thermal stability Water cooled Water cooled Water cooled Water cooled Water cooled Water cooled
Cryogen Use 0.03 L/hr helium 0.06 L/hr helium
Field Strength 1.5 T 1.5 T 1.5T 1.5T 3.0T 1.5T 1.5T 3.0T 3.0T
Strength 30 mT/m 30 mT/m
5 Guass Fringe Field 2.4 m/3.8 m 3.0 m/5.0 m

VIEW OTHER COMPARISON CHARTS

Let’s Discuss Your Needs

Because we carry equipment from all of the leading medical equipment manufacturers and are “vendor neutral,” we can help you find the equipment solution that best meets your needs.