TOPEX - Superficial Radiation Therapy

Superficial Radiation Therapy | Specs

Overview Technical Specs More About Skin Inquire About
Generator:
  • Type- Constant Potential HV, Input Line- 120 - 230 VAC
X-Ray Tube:
  • Metal Ceramic, Water-cooled, Tungsten Target, End grounded
  • Rating= 120kV/10mA, 40kV/30mA, 1200 watts continuous dissipation
  • Focal Spot Size= 1mm x 1mm
Base Unit Assembly:
  • Basic Space Requirements: 29” x 30”
  • Vertical Travel: up to 68”
  • Horizontal Arm Travel: 17” to 61” from center
  • X-Ray Tube Movement: V&H 180 degrees
  • Integrated Modular Components: Input power, HV generator, Heat exchanger
Operator Control Console:
  • Can be located up to 100’ (30 meters) from base unit
  • Service mode for system set-up and calibration - key entry
Two Treatment Techniques:
  • 100kV @ 8mA, 2.1mm Al. HVL or 70kV @ 10mA, 1.3mm Al. HVL
  • X-Ray output is 460 cGy @ 15cm SSD
Automatic Filter Change (Patent Pending):
  • 2.1mm Al. HVL / 1.3 mm Al. HVL / Pb X-Ray Block
Automatic Warm-up Procedures:
  • Automatically activated from the time of last exposure
  • Four pre-programmed sequences
  • Lead blocker automatically placed over x-ray tube support
RAD Check:
  • Direct radiation measurement of output and pre-treatment verification
System Weight:
  • 350 lbs
Accessories:
  • The SRT 100 includes all the various sized Applicators ranging from 1.5cm to 10.0cm, an Applicator Replacement TIP Kit, a Service Spare Fuse/Hardware Kit, a Cable Retractor Installation Kit and a set of Physicist Calibration Fixtures.
Replaceable Safety Contact Shields:
  • Applicator size specific. Visibility of treated area.
Maximum Field Size:
  • 10 cm.
Maximum Energy:
  • 100 kV